MODEL P8/P108 Differential or Absolute Pressure Transducer
The Tavis Model P8 transducer is ideally suited to applications requiring very low
differential pressure measurements. Available in pressure ranges as low as 0.01 psid,
this differential pressure transducer features lowpower consumption and zero and full
scale adjustments. The all welded stainless steel construction and a stable electronics
package combine to produce an extremely accurate, but rugged unit.

HIGH ACCURACY |
LONG TERM STABILITY |
STABLE CALIBRATION |
ALL WELDED CONSTRUCTION |
VERY LOW
PRESSURE RANGES |
|
|
|
GENERAL SPECIFICATIONS |
Pressure Ranges
Proof Pressure
Line Pressure
Weight
Sensor Type
Media |
|
0 to .01 through 0 to 350 psi
20 psid or 200% F.R. (Higher proof pressures available. Consult factory)
20 psia or 200% F.R. (Higher line pressures available. Consult factory)
3.5lbs max.
Variable reluctance
Liquids or gases compatible with 410 SS, 17-4 PH SS and Iconel |
|
|
ELECTRICAL |
Input Voltage
Input Current
Output
Output Impedance
Output Noise
Insulation Resistance
Isolation Resistance |
|
20 to 40 VDC
10 mA max.
0 to 5 VDC
100 ohms max.
15 mV P-P max.
100 megohms min. at 100 VDC
100 megohms min. at 100 VDC |
|
|
PERFORMANCE |
Static Error Band *
Hysteresis
Repeatability
Temperature Error *
Frequency Response *
Regulation
Resolution |
|
± 0.5% F.S. max. (2)
± 0.1% F.S. nom.
± 0.1% F.S. nom.
± 2.0% F.S. max. (2)
Flat ± 5% to 250 Hz
0.05% F.S. per volt max.
Infinite |
|
|
ENVIRONMENTAL |
Temperature Range:
Compensated
Operating
Vibration
Shock
Humidity, Altitude, Rain,
Fungus, Salt spray, Sand,
and Dust |
|
- 65 F to + 165 F (2)
- 65 F to + 200 F
20 g s Peak to 2000 Hz
100 g s, 11 ms
Unaffected
|
*Static Error Band is defined as
the maximum allowed deviation
from a best fit straight line
which minimizes errors due to
the combined effects of nonlinearity,
hysteresis, resolution,
and non-repeatability. |
|
*Temperature Error is defined
as the maximum allowed
deviation from a best fit straight
line which minimizes errors due
to temperature over the range of
65 F to + 165 F. |
|
*Frequency Response given
is for electronics only. Actual
frequency response will depend
on pressure range and media. |
(1) |
All Tavis transducers can be supplied for corrosive media applications.Materials used for this purpose are 17-4 PH SS, 17-7 PH SS and Iconel . Specify model P108 for this feature. |
(2) |
Closer tolerances and higher temperature operation may be supplied. Consult factory for special requirements. |
(3) |
This unit is also available in a two wire 4 to 20 mA configuration. Add C following model number for this feature |
|


"Let us handle your pressure" |